2100P Turbidimeter

Manufacturer: Hach
Order / Quote

The Hach 2100P portable turbidimeter offers a level of performance previously possible only with laboratory instruments. Microprocessor-controlled operation and Hach’s unique Ratio™ optics bring great accuracy, sensitivity, and reliability to field and in-plant testing.

Get fast, accurate turbidity testing in the field or the lab, over a wide range of samples. Compliant with USEPA Method 180.1 design criteria.

The two-detector optical system compensates for color in the sample, light fluctuation, and stray light, enabling analysts to achieve laboratory-grade performance on a wide range of samples, even under difficult, onsite conditions.

Measurement Method

  • Nephelometric ratio

Regulatory

  • Meets EPA Method 180.1

Light Source

  • Tungsten lamp

Range

  • Automatic Range Mode 0 to 1000 NTU
  • Manual Range Selection 0 to 9.99, 0 to 99.9 and 0 to 1000 NTU

Accuracy

  • ±2% of reading plus stray light

Repeatability

  • ±1% of reading, or 0.01 NTU, whichever is greater

Resolution

  • 0.01 on lowest range

Signal Averaging

  • Selectable on/off

Power Requirement

  • 4 AA alkaline batteries or optional battery eliminator

Battery Life

  • Typical 300 tests with signal average mode off
  • 180 tests with signal average mode on

Operating Temperature

  • 0° to 50° C (32° to 122° F)

Sample Required

  • 15 mL (0.5 oz.)

Sample Cells

  • 60 x 25 mm (2.36 x 1 in.) borosilicate glass with screw caps

Dimensions

  • 22.2 x 9.5 x 7.9 cm (8.75" x 3.75" x 3.12")

Weight

  • 0.5 kg (1.1 lb.); shipping weight 2.7 kg (6 lb.)
Specifications subject to change without notice