PortaCount 8020

Manufacturer: TSI Incorporated
Order / Quote

Key Features

  • Computes respirator fit factor using microscopic particles in ambient air
  • Measures fit factors to greater than 10,000
  • Tests any respirator type with a tight fitting facepiece
  • Includes FitPlus™ Fit Test Software for automating the fit test
  • Set pass/fail level from the keypad
  • Set number of exercises from the keypad
  • Digital display indicates the exercise number in progress
  • Direct readout of fit factor
  • Optional sampling adapters allow employees to be tested in their own respirators
  • Compatible with the new N95-Companion for fit testing Series 95 disposable respirators

Software Features

  • OSHA–compliant fit test protocol
  • Pre-loaded for compliance with OSHA, ANSI and international fit test protocols
  • User-definable pass/fail level
  • Real-time fit factor display
  • Search for, retrieve, view on screen and print individual test records
  • Large on-screen exercise descriptions enable you to “oversee” the test from a distance
  • Customize your own report formats
  • Optional card printer

Fit Factor Range

  • 1 to greater than 10,000

Concentration Range

  • 0.01 to 5 × 105 particles/cm3

Particle Size Range

  • 0.02 to greater than 1 micrometer

Typical Fit Factor Accuracy

  • ±10% of reading

Size

  • 9.5 in. × 7.5 in. × 5.5 in. (24 cm × 19 cm × 14 cm)

Weight

  • Unit Only 3.3 lb (1.5 kg)
  • With Standard Accessories 13 lb (5.9 kg) and Case

Power

  • Autosensing 100 to 250 VAC, 50 to 60 Hz

Flow Rate

  • Sample: 100 cm3/min
  • Total: 700 cm3/min (nominal)

Temperature Range

  • Operation: 32 to 100° F (0 to 38° C)
  • Storage: -40 to 160° F (-40 to 70° C)

Alcohol

  • Hours Per Charge: 8 hours at 70° F (21° C)
  • Alcohol Type: 99.5%+ reagent grade isopropyl

RS-232 Output Baud Rate

  • User-selectable: 300, 600, 1200, 2400, 9600 (factory set at 1200)

Carrying Case Size

  • 21 in. × 14 in. × 8.3 in. (53 cm × 36 cm × 21 cm)

Pass/Fail Setting

  • User-selectable: 0, 100, 250, 500, 1000, 1250, 1667, 2000, 4000, 5000, 6667, 10000 (other values can be programmed by user)

Factory Recalibration Interval

  • One year

Download
Download