Manufacturer: RAE Systems
Order / Quote
The IAQRAE is an Indoor Air Quality (IAQ) monitor to combine a Photo-Ionization Detector (PID) with traditional measures of IAQ like Temperature, Humidity, Carbon Dioxide (CO2) and a user-selected toxic gas sensor in one 1-5 sensor package. The PID gives the IAQRAE the ability to measure 0-1000 ppm of VOCs (Volatile Organic Compounds) with 0.01 ppm (10 ppb) resolution.
Dimensions
- 9.25” L x 5.0” W x 9.25” H (23.5cm L x 12.7cm x 23.5cm)
Weight
- 8.9 lbs (4.03 kg) with battery
Detectors Up to five sensors
- PID Sensor w/ 3/8 od 10.6 eV lamp
- Electrochemical Sensor
- NDIR Sensor for CO2
- Humidity Sensor
- Temperature Sensor
Battery
- Rechargeable, 7.4V/7.2 Ah, Lithium-Ion battery pack with built-in charger (less than 10 hours charge time)
- Six C-cell alkaline battery pack
Operating Time
- Up to 36 hours of continuous operation
Display
- 2-line, 16-character LCD
- Manual LED backlight
Keypad
- [MODE] – operation & programming
- [Y/+] – operation & programming key
- [N/-] – operation & programming key
- [RADIO] – not used
- [VOICE] – not used
- [UP/DOWN] – not used
Direct Readout
- Up to five instantaneous values
- Sensor name
- High and low values all sensors
- Battery voltage
- Elapsed time
Alarm Settings
- Separate alarm limit settings for Low and High alarm
- Visible Alarm 100 dB buzzer at 10 cm (typical)
- Audible Alarm Flashing red LED cluster to indicate exceeded preset limits, low battery, or sensor failure
Calibration
- Two point field calibration for fresh air and standard reference gas (except for the temperature sensor)
Attachments
- Shoulder strap, optional tripod/wall mounting bracket
Sampling Pump
- Internal integrated diaphragm pump with adjustable duty cycle ratio of pump on (>400 cc/min) and off
Protection
- Password protected calibration settings, alarm limits, and data
EM Immunity
- No effect when exposed to 0.43 mW/cm2 RF interference (5 watt transmitter at 12 inches)
Data Storage
- Optional 20,000 readings (64 hours, 5 channels at 1-minute intervals) in nonvolatile memory
Datalog Interval
- Programmable 1-3600 seconds
External Alarm
- Optional plug-in, pen-size vibration alarm or earphone
Communication
- Upload data to PC and download monitor setup from PC through RS-232 link to serial port on PC
Temperature
- 0o to 50o C (32o to 122o F)
Humidity
- 0 % to 95% relative humidity (Non-condensing)
Sensors
Sensor | Range | Resolution | t90 |
VOC | 0-9.99 ppm | 0.01 ppm | 10 seconds |
10.0-99.9 ppm | 0.1 ppm | ||
100-999 ppm | 1 ppm | ||
CO2 | 0-20,000 ppm | 10 ppm | 60 seconds |
CO | 0-500 ppm | 1 ppm | 25 seconds |
H2S | 0-100 ppm | 1 ppm | 30 seconds |
SO2 | 0-20.0 ppm | 0.1 ppm | 15 seconds |
NO | 0-250 ppm | 1 ppm | 20 seconds |
NO2 | 0-20.0 ppm | 0.1 ppm | 25 seconds |
Cl2 | 0-10.0 ppm | 0.1 ppm | 60 seconds |
HCN | 0-100 ppm | 1 ppm | 60 seconds |
NH3 | 0-50 ppm | 1 ppm | 150 seconds |
PH3 | 0-5.0 ppm | 0.1 ppm | 60 seconds |
Temp | 32.0° F-122.0° F | 0.1°F | 300 seconds |
0° C-50.0° C | 0.1°C | ||
RH | 5-95% | 1% RH | 60 seconds |