GilAir-5 Air Sampler

Manufacturer: Sensidyne
Order / Quote

The GilAir® pump (GilAir-5) is a reliable 5 liter per minute air sampling pump with automatic constant flow assuring sampling is maintained within ±5% of the initial set point. Setting the flow rate on the GilAir-5 is a simple process that requires no special adjustment tool so you spend less time configuring pump.

Automatic constant flow – the most ­important feature of any sampling pump – is standard on all GilAir-5 pumps. It assures flow is maintained within ±5% of the initial set point, even with varying back-pressures from flow restrictions or buildup of material on the filter. GilAir-5 pump is suitable not only for conventional pull-through sampling media but also for bag sampling and other pressure applications.

The GilAir-5 is now available with nickel-metal hydride (NiMH) batteries that virtually eliminate memory effects and lazy battery issues. NiMH batteries are environmentally friendly, recyclable and do not pose a risk to the environment when being disposed. New NiMH battery packs are backwards compatible with GilAir-5 pumps that currently use NiCd batteries and are user replaceable by removing only two screws on the rear of the pump.

Key Features

Basic, Clock, & Programmable Models The GilAir-5 are available in 3 models; Basic, Clock and Programmable. The Basic model is straight-forward and easy to use. Clock models provide an elapsed-time clock display. Programmable models enable a timer which can start or stop the pump at preset time intervals or program a delayed start.

Flow Fault & Test Indicators

The fault light indicator activates when the pump flow control is outside ±5% of the required flow rate, such as from low battery, blocked tubing or filter. If fault conditions persist for 30-60 seconds, the pump automatically shuts down and freezes the timer. Models with a timer display preserve run time on the display, ensuring a valid sampling result. When the fault condition is cleared before shutdown, the pump will resume normal sampling and fault light turns off. The GilAir-5 also has a Battery Test indicator light that illuminates when the battery can provide a minimum of 8 hours of operation over operating range.

External Filter Housing

Clear filter housing allows an immediate visual inspection of the internal filter’s condition. This feature protects the pump from dirt and debris and is externally mounted on the pump, allowing for quick inspection. Filter replacement requires the removal of just four screws. A built-in moisture trap helps prevent moisture from entering and damaging the pump chamber, such as moisture carryover from impingers.

Optional Low Flow Modules

  • Constant Flow module allows sampling from 20-500 cc/min.
  • Constant Pressure (Multi-flow) Module allows sampling from 1-750 cc/min.
These modules can be quickly added to pumps to adapt them to the lower flow rates required for sorbent tube sampling. The multi-flow module maintains constant pressure and allows sampling with multiple sorbent tubes, each with an independently set flow rate.

Flow Range

  • 1- 5000 cc/min (Total), 850 - 5000 cc/min (High Flow)
  • 5-500 cc/min (Constant Low Flow)
  • 1-750 cc/min (Constant Pressure)

Constant Flow Control

  • ± 5% of set flow at 1000-5000 cc/min

Flow Compensation

  • 5000cc up to 8” H2O back pressure
  • 4000cc up to 18” H2O back pressure
  • 3000cc up to 23” H2O back pressure
  • 850cc up to 29” H2O back pressure

Run Time

  • 8 hour minimum

Flow Fault

  • If flow changes exceed 5%, fault icon appears
  • If fault exceeds 30 seconds, pump shuts down
  • Pump attempts to restart every 3 minutes for up to 1 hour

Temperature Ranges

  • Operating: 32°F to 104°F (0°C to 40°C)
  • Storage: -49°F to 113°F (-45°C to 45°C)
  • Charging: 41°F to 104°F (5°C to 40°C)

Humidity Ranges

  • Operating: 0-85 %RH, non-condensing
  • Storage: 0-100 %RH, non-condensing

Display Messages

  • LAST, CAL, SHUT/OFF (Clock, Program); E, PC, P1-P6, OLD (Program)

Controls

  • Power Switch, Flow Control Screw & Buttons (Clock, Program), Programming (PROG) Button (Program)

Indicators

  • Elapsed Time (Clock, Program)

Icons (LCD)

  • Low Battery, Flow Fault, Clock (Clock & Program)

Dimensions

  • 3.9W x 4.1H x 2.0D inches 10.0W x 9.0H x 5.1D cm

Weight

  • 22.5 oz. (638 g)

Battery Pack

  • UL: NiCad or NiMH
  • ATEX: NiMH

Interface Connectors

  • Charging Jack

Charging Time

  • 14-18 hours

CE

  • EMC EMI/RFI, EN 55 022 Class B; IEC 801-2, 3

Intrinsic Safety

  • UL: Class I, Div 1, Groups A, B, C, D; T3C
  • CENELEC: EEx ia IIC T4 (ATEX models)

Download